Mycophenolic Acid (MPA) Level
Collection Requirements
Source: blood
Container: Red
Volume: 2 mL in vacutainer
Special Instructions
- Send to the lab immediately following collection
- Transplant patient kinetics studies – all 4 collection timepoints must be received in the lab before testing will be done
Turn Around Time
1-4 days
Testing performed Monday, Wednesday, and Friday
Day of testing batch cut offs:
- Transplant patient kinetics studies: all 4 collection timepoints must be received in the Toxicology Lab by 2 p.m. to be included in the batch.
- All other collections: must be received in the Toxicology Lab by 11 a.m. to be included in the batch.
Availability
Routine
Lab Processing Instructions
Specimen: serum
Spin, separate, and refrigerate
Note: serum must be separated within 2 hours of collection
Performing Laboratory
Adele Hall
Adele Hall Lab Section
Biochemical Genetics
Instrumentation/Methodology
High Performance Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS)
Reference Ranges
Mycophenolic Acid (MPA) 1.0-3.5 mcg/mL
MPA Glucuronide 35-100 mcg/mL
CPT
80180